  • The Z5+ probe measures X, Y and Z heights for 5 separate laser spots simultaneously.  While the center spot in the pattern is stationary, the other four can be manually positioned within the field of view.

  • A process vision camera provides feedback to the user for easy positioning of the measurement locations and verifying them on the measured product.

 

 

Features


Preliminary Performance Specifications

  X Y Z
Measurement range ±0.75 mm ±0.75 mm ±1.25 mm1
Resolution (N=1) 0.11 µm 0.11 µm 0.11 µm
Resolution (N=5) 0.08 µm 0.08 µm 0.08 µm
RMS Repeatability 0.11 µm 0.11 µm 0.11 µm

Angular Resolution2 (N=1) 1 arc-min
Thermal Stability <0.002%F.S./°C
Laser Wavelength / Power 660 nm / < 12mW
Sensor Technology 1/3" CCD (x3)
Standoff 17mm
Spot Size ~50 µm
Measurement Speed ~150 msec
Physical Dimensions (mm) 180 (H) x 160 (W) x 80 (D)
Power Requirements None (supplied via cameras)
  1. Effective range for all spots when in 1mm square pattern. Range of >±1.5mm can be achieved on smaller patterns.
  2. Numerically deduced value based on geometires for 0.5mm square spot pattern.

Software

Setup Program
Each probe comes with a simple yet powerful sample program which is very useful for setup, troubleshooting, testing and learning.

Custom Program
AKI can also provide a custom program tailored to the customer’s exact needs.  This can include interfacing with other hardware and/or motion control.

Integration into your own program. 
Developers can choose to write their own program using the AKI Software Libraries. You may choose from either an ActiveX COM object or a .NET dll in order to communicate with the AKI Probe. Integration is extremely simple and source code examples are supplied for any development language.

  • Visual C++ 6.0 MFC
  • Visual Basic 6.0
  • C#.NET VS 2003/2005
  • VB.NET VS 2003/2005
  • C++.NET VS 2003/2005
  • Labview 7.1/8.0

*Software is based on Cognex Vision Pro Libraries

  • Win2000
  • WinNT4 SP6
  • WinXP (SP1, SP2)

Z5+ Mounting Diagram

Notes:

  1. 5K7 holes are press fit for M5 dowel pins.
  2. Dowels to protrude 5mm max.
  3. Optionally, the M5 tapped holes may be made as clearance holes for M6 screws for screw mounting from the rear of the mount plate.
Documents
AKI OCX Software Reference.pdf - Software Command Reference Rev 1-19-07
Z5+ Information - Z5+ Information
 